Ultrasound Transducer Leakage Current Meter

SECULIFE UL

 

IEC 61010, IEC 60335, IEC 60950, VDE 0701-0702

The SECULIFE UL is specifically designed for electrical safety testing of all types of diagnostic ultrasound transducers, independent of the ultrasound equipment typically used.

The SECULIFE UL can be applied to all types of ultrasound transducers, particularly Transesophageal Echocardiography (TEE) probes. According to IAC standards, leakage current testing must be performed on all TEE probes before each use.

The SECULIFE UL detects insulation integrity and conducts capacitive leakage current tests. It is the most advanced instrument of its kind on the market, ensuring the electrical safety of ultrasound transducers.

Features:

  • Editable upper and lower leakage current limits with PASS/FAIL indication
  • Measurement range: 0.5 to 500 µA
  • 1% measurement accuracy
  • Selectable test voltage/frequency: 90~275V, 50/60Hz
  • 99 data storage slots
  • RS232 communication interface for computer connection, enabling editing and remote control
  • Optional external printer available
 

Specifications

Model
SECULIFE UL
-